NELSON ANDRES, Valdez Velázquez y RAFAEL, Klassen Esau. Quantitative and Qualitative Evaluation of Noise in Magnetic Resonance Imaging. Measured Values, ISO 9612 Standards, and Biological Effects. An. Fac. Cienc. Méd. (Asunción) [online]. 2026, vol.59, n.1, e59012617. Epub 30-Abr-2026. ISSN 1816-8949. https://doi.org/10.18004/anales/2026.e59012617.
This study focuses on the evaluation of magnetic noise based on sound pressure levels and its impact on the health of exposed personnel and patients in a Magnetic Resonance Imaging Service in the city of Asunción.
Objectives:
Application of formulas to sound pressure values measured with a sound level meter, according to the analytical model of the ISO-9612 standard. Additionally, scientific databases were consulted in order to identify possible biological effects.
Materials:
Magnetic Resonance Imaging equipment General Electric, Signa Prime 1.5T AIR Recon DL; sound level meter Quest Technologies 1900 e; sound level meter PCE-322A; ISO-9612 analytical model.
Methods:
In situ noise level measurements were carried out; the Task-Based Measurement Strategy of ISO 9612 was applied to calculate the daily weighted 8-hour acoustic exposure of clinical personnel; and finally, an extensive literature review was conducted to identify possible biological effects of noise exposure on personnel and patients.
Results:
Noise levels within the equipment room reached peaks of up to 93 dBA, while measurements in the control room remained within acceptable limits. The calculated sound exposure level indicated that the occupational risk of chronic hearing loss is classified as low. Finally, the observed biological effects are consistent with the measured noise levels. There are effects caused by the intensity of acoustic stress and other possible reported effects that cannot be ruled out if protective measures fail.
Palabras clave : Magnetic Resonance Imaging; Noise; Noise Control; Hearing Threshold; Occupational Exposure; Physiological Phenomena; ISO 9612..